Quick
Search: 
 
advanced search
 GSW Home    GeoRef Home    My GSW Alerts    Contact GSW    About GSW    Journals List    Help 
American Mineralogist RIMG advertisement
JOURNAL HOME HELP CONTACT PUBLISHER SUBSCRIBE ARCHIVE SEARCH TABLE OF CONTENTS

American Mineralogist; January 2006; v. 91; no. 1; p. 111-114; DOI: 10.2138/am.2006.1883
© 2006 Mineralogical Society of America
This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Web of Science (1)
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Nishihara, Y.
Right arrow Articles by Karato, S.-I.
Right arrow Search for Related Content
GeoRef
Right arrow GeoRef Citation

Effects of metal protection coils on thermocouple EMF in multi-anvil high-pressure experiments

Yu Nishihara1,*, Kyoko N. Matsukage1,2 and Shun-Ichiro Karato1

1 Department of Geology and Geophysics, Yale University, New Haven, Connecticut 06520, U.S.A.
2 Department of Environmental Sciences, Ibaraki University, Bunkyo, Mito, Ibaraki 310-0056, Japan

Correspondence: * Corresponding author present address: Department of Earth and Planetary Sciences, Tokyo Institute of Technology, Meguro-ku, Tokyo 152-8551, Japan. E-mail: yuu{at}geo.titech.ac.jp

Metallic coils (in most cases Cu coils) are often used in high-pressure experiments to protect thermocouple wires. In this paper we show that these coils have important influences on thermocouple EMF and therefore on the temperature measurements. We tested this effect by measuring EMF from Cu coiled single wires of chromel and alumel, and, further, we conducted experiments to compare the EMF from W5Re-W26Re thermocouples with and without Cu coils attached to them. The results show systematic differences in thermocouple readings; the EMFs from W-Re thermocouples with Cu coils give systematically lower values than EMFs from thermocouples without Cu coils. The results were analyzed using a simple model. The difference in thermocouple EMFs between thermocouples with and without protection coils is given by


Formula

where Ei and Ri are the EMF and the electrical resistance of metal i in the portion of the Cu coil, and the subscripts 1+, 1–, and 2 indicate positive thermocouple metal, negative thermocouple metal, and coil metal, respectively. The addition of a coil with different metal has a large effect — the {Delta}ETC will be close to –(E1+E1–) < 0 when the resistance of the coil is significantly smaller than that of a thermocouple wire. For a Cu coil and W-Re thermocouple, R1+, 1– >> R2 and therefore thermocouple readings with a Cu coil will lead to underestimation of the real temperature. Under common experimental conditions with a multi-anvil apparatus, the error in the temperature estimate caused by Cu protection coils is ~100–150 K for a peak temperature of 1600–2000 K.

Key Words: High-pressure and high-temperature studies • thermocouple EMF • electrical properties • tungsten-rhenium alloy • chromel-alumel




This article has been cited by other articles:


Home page
American MineralogistHome page
A. Watenphul, B. Wunder, and W. Heinrich
High-pressure ammonium-bearing silicates: Implications for nitrogen and hydrogen storage in the Earth's mantle
American Mineralogist, February 1, 2009; 94(2-3): 283 - 292.
[Abstract] [Full Text] [PDF]


Home page
American MineralogistHome page
Y. Nishihara, T. Shinmei, and S.-i. Karato
Effect of chemical environment on the hydrogen-related defect chemistry in wadsleyite
American Mineralogist, May 1, 2008; 93(5-6): 831 - 843.